Product Overview
Portable MDM300 / MDM300 I.S. dew-point hygrometer for rapid spot checks in compressed air and natural gas. Request price, availability and a quotation for MDM300.
Key Parameters
- Dew-point accuracy +/-1 deg C from -60 to +20 deg Cdp
- MDM300: T95 to -60 deg Cdp in up to 15 minutes
- MDM300 I.S.: T95 to -60 deg Cdp in up to 30 minutes
- Operating pressure up to 350 barg
Highlights
- MDM300: up to 48 hours of typical use
- MDM300 I.S.: up to 24 hours of typical use
- IP66 / NEMA 4 portable enclosure
- Data logging and configurable sampling accessories
Applications
- Natural gas processing, pipelines and dew-point spot checks
- Monitoring desiccant dryers for compressed air
- Petrochemical, industrial gas and medical gas applications
What to specify for selection
- Gas type and measurement units
- Working pressure and expected dew-point range
- The standard MDM300 is for safe-area use. For a hazardous area, consider MDM300 I.S. only after checking the site's area classification and the specific unit's approval conditions. Confirm gas compatibility and sampling requirements for the intended application.
- Sample connection, sampling accessories and quantity